AMX Wave Wireless Accessories and Adapters none Manual de usuario Pagina 20

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 44
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 19
20
Rosemount 3300 Series
October 2014
www.rosemount.com
Performance specifications
General
Reference Conditions Twin Lead probe, 77 °F (25 °C) water
Reference Accuracy
± 0.2 in. (5 mm) for probes 16.4 ft. (5 m)
± 0.1% of measured distance for rigid probes >16.4 ft. (5 m)
± 0.15% of measured distance for flexible probes >16.4 ft. (5 m)
Repeatability ± 0.04 in. (1 mm)
Ambient Temperature Effect Less than 0.01% of measured distance per °C
Update Interval 1 per second
Measuring range
Transition Zones
These zones are areas where
measurements are non-linear or have
reduced accuracy. If measurements are
desired at the very top of a tank, it is
possible to mechanically extend the
nozzle and use a coaxial probe. The
upper transition zone is then moved
into the extension. See Table 5 on
page 22.
Measuring Range and
Minimum Dielectric Constant
16 in. (0.4 m) to 77 ft. (23.5 m)
See Table 6 on page 22 for each probe’s measuring range and minimum dielectric constant. Due to
the measuring range depending on the application and factors described below, the values are a
guideline for clean liquids. For more information, ask your local Emerson Process Management
representative.
Different parameters (factors) affect the echo and therefore the maximum measuring range differs
depending on application according to:
Disturbing objects close to the probe
Media with higher dielectric constants (
r
) give better reflection and allow a longer measuring
range
Surface foam and particles in the tank atmosphere may affect measuring performance
Heavy coating or contamination on the probe should be avoided since it can reduce measuring
range and might cause erroneous level readings
Note
See Table 7 on page 23 for the measuring range when using the Remote Housing.
Upper Reference Point
Upper Transition Zone
Maximum
Recommended
Measuring Range
Lower Transition Zone
Lower Reference Point
Lower Transition Zone
For a flexible single lead probe with chuck,
the lower transition zone is measured
upwards from the upper part of the clamp.
Lower Transition Zone
Vista de pagina 19
1 2 ... 15 16 17 18 19 20 21 22 23 24 25 ... 43 44

Comentarios a estos manuales

Sin comentarios